Gamme de fréquence: 9KHz ~ 3.2GHz ;
Gamme de generador de Tracking: 5M~1.6GHz TG
Sensibilité optimale: -161dBm ;
RBW: 10 Hz à 1 MHz.
Design portable haute protection, robuste et durable.
Peut être utilisé en laboratoire, plus adapté pour une utilisation mobile sur le terrain.
Batterie au lithium de grande capacité, peut fonctionner pendant plus de 4 heures. 18650 batterie rechargeable.
Communication USB / LAN, facile pour le contrôle à longue distance.
Écran couleur de 5,6 pouces, résolution 640 * 480, affichage de la forme d'onde clairement.
Features
HSA2000 Series measurement application:
Parameters |
|||
Model |
HSA2030A |
HSA2030B |
|
Frequency |
|||
Frequency Range |
9KHz~3.2GHz AC Coupled |
9KHz~3.2GHz AC Coupled 5M~3.2GHz TG |
|
Frequency Resolution |
1Hz |
||
Reference Frequency |
10MHz |
||
Frequency Readout Accuracy |
±(frequency indication*frequency reference uncertainty+1%*span+20%RBW+marker resolution+1Hz) |
||
Internal 10MHz Reference Accuracy |
Aging rate |
±1ppm/year ( 0 ℃ ~50 ℃ , Reference is 25 ℃ ) |
|
Temperature stability |
±1ppm/year |
||
Marker Resolution |
(Frequency span)/(number of sweep points-1) |
||
Resolution Bandwidth ( RBW ) |
|||
-3dB Bandwidth |
10Hz to 1MHz , 1-3-10 sequence |
||
Accuracy |
±5% RBW=10Hz~1MHz nominal |
||
Resolution Filter Shape Factor |
< 5:1 nominal |
||
Video Bandwidth ( VBW ) |
-3dB bandwidth |
1Hz to 1MHz , 1-3-10 sequence |
|
Accuracy |
±10% VBM=1Hz~1MHz nominal |
||
Displayed Average Noise Level (normalized to1Hz) |
|||
9K~1MHz |
Preamp off |
-108dBm , typical -127dBm |
|
1MHz ~10MHz |
-128dBm , typical -146dBm |
||
10MHz ~500MHz |
-142dBm , typical -146dBm |
||
500MHz ~2.5GHz |
-141dBm , typical -145dBm |
||
2.5GHz ~3.2GHz |
-140dBm , typical -144dBm |
||
9K~1MHz |
Preamp on |
-131dBm , typical -150dBm |
|
1MHz ~10MHz |
-148dBm , typical -163dBm |
||
10MHz ~500MHz |
-161dBm , typical -164dBm |
||
500MHz ~2.5GHz |
-159dBm , typical -162dBm |
||
2.5GHz ~3.2GHz |
-158dBm , typical -161dBm |
||
SSB Phase Noise |
|||
Carrier Offset(20 ℃ ~ 30 ℃, 500MHz Central Frequency ) |
10K |
< -92 dBc/Hz, typical -95 dBc/Hz |
|
30K |
< -93 dBc/Hz, typical -96 dBc/Hz |
||
100K |
< -95 dBc/Hz, typical -97 dBc/Hz |
||
1MHz |
< -117 dBc/Hz, typical -119 dBc/Hz |
||
Sweep Time |
|||
Range |
Span >100Hz |
2ms to 1000s |
|
Span =0Hz |
600ns to 200s |
||
Sweep Mode |
Continuous, single |
||
Trigger Source |
Free run; video; external |
||
Trigger slope |
Selectable positive or negative edge |
||
Trigger delay |
Span =0Hz |
±12ms to ±12s nominal |
|
Frequency Counter |
|||
Counter Resolution |
1Hz |
||
Accuracy |
± (marker frequency × frequency reference uncertainty + counter resolution) |
||
Level Display Range |
|||
Log Scale and Units |
1 to 10 dB/divisions in 1, 2, 5, 10 dB steps, 10 divisions displayed |
||
Linear Scale and Units |
0 to 100% , 10 divisions displayed |
||
Scale Unit |
dBm , dBmV,dBuV,Watts , Volts |
||
Sweep (Trace) Points |
461 |
||
Number of Markers |
4 |
||
Detectors |
Normal, positive peak, dample, negative peak, RMS |
||
Number of Traces |
4 |
||
Trace Functions |
Clear/write, maximum hold, minimum hold, average, check , close |
||
Level Measurement Error |
± 1.5dB(excluding input VSWR mismatch) 20~30 ℃, peak detector, preamplifier off, input signal -50dBm to 0dBm |
||
Reference Level |
|||
Setting Range |
-100dBm to +30dBm, steps of 1dB |
||
Setting Resolution |
Log scale |
0.01dB |
|
Linear scale |
Almost log ( 2.236μV to 7.07V ) |
||
Amplitude |
|||
Maximum Safety Input Level |
Average continuous power |
+33dBm |
|
DC input voltage |
50Vdc |
||
Measurement Range |
9KHz~2MHz |
Displayed average noise level (DANL) to +10dB |
|
2MHz~3.2GHz |
Displayed average noise level (DANL) to +20dB |
||
Input attenuator range |
0 to 51dB,1dB steps |
||
Spurious Response |
|||
Second Harmonic Distortion ( SHI ) |
<65dBc , 50MHz to 3.2GHz ( Mixer level -30dBm , attenuator =0dB , preamp off , 20 ℃ ~30 ℃ ) |
||
Third-Order Intermodulation ( TOI ) |
50~300MHz |
+8dBm , Third-order intermodulation products : 2 x -20dBm; frequency separation 100KHz : attenuation = 0dB; preamp off , 20 ℃ ~30 ℃ |
|
300MHz~3.2GHz |
+10dBm |
||
Input Related Spurious |
<-75dBc , (input mixer = -30dBm) |
||
Inherent Residual Response |
<-90dBm , typical -98dBm (Input terminated and 0 dB RF attenuation, preamplifier off) |
||
RF Input VSWR (at Tuned Frequency) |
10MHz to 3.2GHz |
< 1.5:1, nominal Attenuator setting 10~20dB |
|
10MHz Reference/External Trigger Input |
|||
Reference Iinput Frequency |
10MHz |
||
Reference Input Amplitude |
0~10dBm |
||
Trigger Voltage |
5V TTL level |
||
Connector and Output Impendence |
N female ;( 50Ω ) |
||
Gennral Feature |
|||
Interface Language |
English , Chinese , Chinese Traditional |
||
Display Index |
5.7 inch , 640*480 resolution , 64M color LCD display |
||
Temperature Range |
Working |
-10 ℃ to +50 ℃ ,( battery : 0 ℃ to 50 ℃ ) |
|
Storage |
-40 ℃ to +70 ℃ ,( battery :-20 ℃ to 50 ℃ ) |
||
Relative Humidity |
<95% |
||
Weight |
2.9kg ( with battery ), 2.6kg ( without battery ) |
||
Size |
260m X 220m X 75m |
||
Power |
Input voltage range |
DC: 12-17V, maximum 2.8A input 220VAC±15% |
|
AC frequency range |
40Hz to 60Hz |
||
Power consumption |
Maximum 32 W |
||